Sign in →

Test Code MISCLM - LAB2608 Neuron-Specific Enolase, Serum

Important Note

GOKEY: NSE - Neuron Specific Enolase, S

Additional Codes

SQ: MISCLM

MAYO: NSE

Reporting Name

Neuron Specific Enolase, S

Useful For

A follow-up marker in patients with neuron-specific enolase-secreting tumors of any type

 

An auxiliary test in the diagnosis of small cell lung carcinoma

 

An auxiliary test in the diagnosis of carcinoids, islet cell tumors, and neuroblastomas

 

An auxiliary tool in the assessment of comatose patients

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Specimen Required


Collection Container/Tube:

Preferred: Red top

Acceptable: Serum gel

Submission Container/Tube: Plastic vial

Specimen Volume: 0.5 mL

Collection Instructions:

1. Specimens should not be transported by tube system prior to centrifugation.

2. Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.3 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 7 days
  Ambient  5 days

Reference Values

≤15 ng/mL

Serum markers are not specific for malignancy, and values may vary by method.

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

83520

LOINC Code Information

Test ID Test Order Name Order LOINC Value
NSE Neuron Specific Enolase, S 15060-7

 

Result ID Test Result Name Result LOINC Value
NSE Neuron Specific Enolase, S 15060-7

Report Available

1 to 3 days

Reject Due To

Gross hemolysis Reject
Gross lipemia OK
Gross icterus Reject
Hemolysis at any level Reject

Day(s) Performed

Monday through Saturday

Method Description

Neuron-specific enolase is measured in this homogeneous automated immunofluorescent assay on the BRAHMS Kryptor. The Kryptor uses TRACE (time resolved amplified cryptate emission) technology based on a nonradioactive transfer of energy. This transfer occurs between 2 fluorescent tracers: the donor (europium cryptate) and the acceptor (XL665). In the NSE assay, 2 monoclonal antibodies are labeled, 1 with europium cryptate and 1 with XL665. NSE is sandwiched between the 2 antibodies, bringing them into close proximity. When the antigen-antibody complex is excited with a nitrogen laser at 337 nm, some fluorescent energy is emitted at 620 nm, and the rest is transferred to XL665. This energy is then emitted as fluorescence at 665 nm. A ratio of the energy emitted at 665 nm to that emitted at 620 nm (internal reference) is calculated for each sample. Signal intensity is proportional to the number of antigen-antibody complexes formed and, therefore, to antigen concentration.(Unpublished Mayo method)

Method Name

Homogeneous Time-Resolved Fluorescence

Forms

If not ordering electronically, complete, print, and send an Oncology Test Request (T729) with the specimen.

Secondary ID

80913