Test Code Epic LAB765 Viscosity, Serum
Additional Codes
SUNQUEST: VISCOM
MAYO: SVISC
Specimen Required
Collection Container/Tube: Red top (serum gel/SST are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 3 mL
Collection Instructions:
1. Keep specimen at 37° C (eg, 37° C Thermopak, heat block) until after centrifugation.
2. Centrifuge and aliquot serum into plastic vial.
Secondary ID
610406Useful For
Detection of increased viscosity
Monitoring patients with hyperviscosity syndrome
This test is not useful for patients with small concentrations of monoclonal proteins.
Method Name
Capillary Measurement
Reporting Name
Viscosity, SSpecimen Type
Serum RedSpecimen Minimum Volume
0.75 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum Red | Refrigerated (preferred) | 28 days | |
Frozen | 28 days | ||
Ambient | 14 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
≤1.5 centipoises
Method Description
The Benson BV200 automated viscometer employs vacuum pressure through a capillary to measure the dynamic viscosity of serum samples. Samples at ambient temperature are aspirated by the sample probe, which is heated to 37° C. Viscosity results are calculated by the instrument at both 25° C and 37° C temperature points. Viscosity units on board the instrument are mPa.s (milli-pascal seconds), which is the International Standard (SI) unit for viscosity. Centipoise (cP) units are named after French physicist Jean Leonard Marie Poiseuille and are used in the centimeter-gram-second (CGS) system of units. 1 mPa.s is equal to 1 cP.(Instruction manual: BV200 Automated Clinical Viscometer User Manual. Benson Viscometers; Ver. 1.5, 02/2020)
Day(s) Performed
Monday through Friday
Report Available
1 to 3 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
85810
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
SVISC | Viscosity, S | 3128-6 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
SVISC | Viscosity, S | 3128-6 |
Forms
If not ordering electronically, complete, print, and send 1 of the following forms with the specimen: