Test Code Epic LAB468 Diazepam and Nordiazepam, Serum
Additional Codes
SQ code DIAZPM
Mayo code DIA
Test Down Notes
Effective December 14, 2023: This test is temporarily unavailable due to analytic issues. The downtime is expected to be >30 days. Order ZW86 (1501SP) as an alternate. See test notification here.
Reporting Name
Diazepam and Nordiazepam, SSpecimen Type
Serum RedPerforming Laboratory

Useful For
Assessing compliance
Monitoring for appropriate therapeutic level
Assessing diazepam toxicity
Specimen Required
Collection Container/Tube: Red top (Serum gel/SST are not acceptable)
Submission Container/Tube: Plastic vial
Specimen Volume: 0.5 mL
Collection Instructions: Centrifuge and aliquot serum into plastic vial.
Specimen Minimum Volume
0.3 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Serum Red | Refrigerated (preferred) | 14 days |
Ambient | 14 days | |
Frozen | 14 days |
Reference Values
Therapeutic concentrations
Diazepam and Nordiazepam: 200-2,500 ng/mL
Day(s) Performed
Monday, Wednesday
Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
80299
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
DIA | Diazepam and Nordiazepam, S | 49044-1 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
8629 | Diazepam | 3548-5 |
2475 | Nordiazepam | 3537-8 |
2459 | Diazepam and Nordiazepam | 16757-7 |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Method Description
The internal standard mixture containing chlordiazepoxide-d5, diazepam-d4, and nordiazepam-d5 is added to serum samples. The serum samples are treated with phosphate buffer and extracted via liquid/liquid extraction. The organic layer from the extraction is dried under nitrogen, reconstituted, and injected on a liquid chromatography tandem mass spectrometer.(Unpublished Mayo method)
Report Available
3 to 7 daysForms
If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.