Sign in →

Test Code Epic LAB4312 Lacosamide, Serum

Additional Codes

SQ code LACO

Mayo code LACO

 

Reporting Name

Lacosamide, S

Specimen Type

Serum

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Useful For

Monitoring serum concentrations of lacosamide to ensure compliance and appropriate dosing in specific clinical conditions (ie, severe kidney impairment, mild-to-moderate hepatic impairment, and kidney failure)


Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Container/Tube:

Preferred: Red top

Acceptable: Serum gel

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions:

1. Collect specimen immediately before next scheduled dose.

2. For sustained-release formulations ONLY, collect specimen a minimum of 12 hours after last dose.

3. Centrifuge and aliquot serum into plastic vial within 2 hours of collection.


Specimen Minimum Volume

0.2 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 28 days
  Ambient  28 days
  Frozen  28 days

Reference Values

Patients receiving therapeutic doses usually have lacosamide concentrations of 1.0 to 10.0 mcg/mL.

Day(s) Performed

Monday through Saturday

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

80235

LOINC Code Information

Test ID Test Order Name Order LOINC Value
LACO Lacosamide, S 59297-2

 

Result ID Test Result Name Result LOINC Value
62772 Lacosamide, S 59297-2

Reject Due To

Gross hemolysis OK
Gross lipemia OK
Gross icterus OK

Method Name

Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS)

Portions of this test are covered by patents held by Quest Diagnostics

Method Description

Lacosamide and the internal standard are separated from other serum constituents by high-performance liquid chromatography with analysis on a tandem mass spectrometer equipped with an electrospray ion source using multiple reaction monitoring.(Unpublished Mayo method)

Report Available

Same day/1 to 4 days

Forms

If not ordering electronically, complete, print, and send one of the following with the specimen:

-Neurology Specialty Testing Client Test Request (T732)

-Therapeutics Test Request (T831)

Secondary ID

62905