Test Code Epic LAB3545 3-Hydroxy-3-Methylglutaryl Coenzyme-A (HMG-CoA) Reductase, Serum
Additional Codes
SQ: HMGCRM
MAYO: HMGCR
Ordering Guidance
NMS1 / Necrotizing Myopathy Evaluation, Serum is the preferred first tier test for identification of antibodies specific for necrotizing autoimmune myopathy (HMGCOA-IgG and SRP-IgG). This initial evaluation includes signal recognition particle (SRP) antibodies performed using tissue indirect immunofluorescence, which increases the clinical sensitivity as compared to SRP immunoblot methodologies.
Specimen Required
Collection Container/Tube:
Preferred: Red top
Acceptable: Serum gel
Submission Container/Tube: Plastic vial
Specimen Volume: 2 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Secondary ID
607414Useful For
Evaluating patients with suspected necrotizing autoimmune myopathy
Measuring 3-hydroxy-3-methylglutaryl-CoA reductase antibodies
Method Name
Chemiluminescence Immunoassay (CIA)
Reporting Name
HMG-CoA Reductase Ab, SSpecimen Type
SerumSpecimen Minimum Volume
1 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 28 days | |
Frozen | 28 days | ||
Ambient | 72 hours |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Gross icterus | Reject |
Reference Values
<20.0 CU
Method Description
IgG antibodies to 3-hydroxy-3- methylglutaryl coenzyme A reductase (HMGCR) are detected by a chemiluminescent assay using the Inova BIO-FLASH instrument. HMGCR antigen is coated on to paramagnetic beads, which are stored in the reagent cartridge lyophilized. When the assay cartridge is ready to be used for the first time, a buffer solution is added to the tube containing the beads, and the beads are resuspended with the buffer. The reagent cartridge is then loaded onto the BIO-FLASH instrument. A patient serum sample is diluted 1:17 by the instrument in a disposable plastic cuvette. An aliquot of the diluted patient serum, HMGCR-coupled beads, and assay buffer are combined into a second cuvette and mixed. This cuvette is incubated at 37° C. The beads are then magnetized and washed several times. Isoluminol conjugated anti-human IgG antibody is then added to the cuvette and incubated at 37° C. Again, the beads are magnetized and washed repeatedly. The isoluminol conjugate produces a luminescent reaction when "trigger" reagents are added to the cuvette. The light produced from this reaction is measured as relative light units (RLU) by the BIO-FLASH optical system. RLU values are proportional to the amount of bound isoluminol conjugate, which in turn is proportional to the amount of anti-HMGCR antibodies bound to the antigen on the beads. The QUANTA Flash HMGCR assay utilizes a predefined lot specific master curve that is uploaded into the instrument through the reagent cartridge barcode. Based on the results obtained by running two calibrators, an instrument specific working curve is created, which is used by the software to calculate chemiluminescent units from the RLU value obtained for each sample.(Package insert: QUANTA Flash HMGCR 701333. Inova Diagnostics, Inc; V04, 09/2018)
Day(s) Performed
Monday through Friday
Report Available
2 to 4 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
82397
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
HMGCR | HMG-CoA Reductase Ab, S | 93493-5 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
607414 | HMG-CoA Reductase Ab, S | 93493-5 |