Test Code Epic LAB2744 Toxoplasma gondii Antibody, IgM and IgG, Serum
Additional Codes
SQ code TOXOM
Mayo code TXMGP
Reporting Name
Toxoplasma Ab, IgM and IgG, SSpecimen Type
SerumPerforming Laboratory
Mayo Clinic Laboratories in RochesterUseful For
Determining whether a patient has had previous exposure to or recent infection with Toxoplasma gondii
IgG is not useful for diagnosing infection in infants younger than 6 months of age. IgG antibodies in this age group usually are the result of passive transfer from the mother.
Testing Algorithm
Specimen Required
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Aliquot tube
Specimen Volume: 1.5 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Specimen Minimum Volume
0.8 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 14 days | |
Frozen | 14 days |
Special Instructions
Reference Values
Toxoplasma IgM
Negative
Toxoplasma IgG
Negative
Toxoplasma IgG Value
≤9 IU/mL (Negative)
10-11 IU/mL (Equivocal)
≥12 IU/mL (Positive)
Reference values apply to all ages.
Day(s) Performed
Monday through Saturday
Test Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
86778-Toxoplasma IgM
86777-Toxoplasma IgG
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
TXMGP | Toxoplasma Ab, IgM and IgG, S | 88746-3 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
TOXG | Toxoplasma Ab, IgG, S | 40677-7 |
TXM | Toxoplasma Ab, IgM, S | 40678-5 |
DEXG6 | Toxoplasma IgG Value | 8039-0 |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Heat-inactivated specimen | Reject |
Method Name
Multiplex Flow Immunoassay (MFI)
Secondary ID
39857Profile Information
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
TXM | Toxoplasma Ab, IgM, S | Yes | Yes |
TOXGP | Toxoplasma Ab, IgG, S | Yes | Yes |
Method Description
The BioPlex 2200 Toxoplasma gondii IgM and IgG assays use multiplex flow immunoassay technology. Briefly, Toxoplasma antigen-coated fluorescent beads are mixed with an aliquot of patient sample and sample diluent and then incubated at 37° C. During this time, IgM and IgG anti-Toxoplasma antibodies in the specimen will bind to the Toxoplasma antigen on the beads. After a wash cycle, a fluorescently-labeled antihuman IgM- and IgG-antibody conjugate is added to the mixture and incubated at 37° C. Following a wash step to remove unbound conjugate, the bead mixture is passed through a detector that identifies the bead based on dye fluorescence and determines the amount of antibody captured by the antigen based on fluorescence of the antihuman-IgG conjugate. Raw data is calculated in relative fluorescence intensity and is converted to an antibody index for interpretation for IgM and an IU/mL for IgG.
Three additional dyed beads: an internal standard bead, a serum verification bead, and a reagent black bead are present in each reaction mixture to verify detector response, the addition of serum to the reaction vessel, and the absence of significant nonspecific binding in serum, respectively.(Package inserts: BioPlex 2200 System, ToRC IgG and ToRC IgM. Bio-Rad Laboratories; 03/2012 and 08/2017)
Report Available
Same day/1 to 3 daysForms
If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.