Test Code Epic LAB2437 Enterovirus, Molecular Detection, PCR, Plasma
Additional Codes
SQ code EVPCRM
Mayo code ENTP
Reporting Name
Enterovirus PCR, PSpecimen Type
Plasma EDTAPerforming Laboratory
Mayo Clinic Laboratories in RochesterUseful For
Aiding in diagnosing enterovirus infections using plasma specimens
This test should not be used to screen asymptomatic patients.
Ordering Guidance
This test will detect enterovirus but will not differentiate viruses in this family or provide serotyping information.
Specimen Required
Submit a raw clinical sample (not a culture isolate) for enterovirus polymerase chain reaction .
Collection Container/Tube: Lavender top (EDTA)
Submission Container/Tube: Screw-capped, sterile container
Specimen Volume: 1 mL
Collection Instructions: Centrifuge blood collection tube and aliquot plasma into plastic vial.
Specimen Minimum Volume
0.3 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma EDTA | Refrigerated (preferred) | 7 days | |
Frozen | 7 days |
Reference Values
Negative
Day(s) Performed
Monday through Sunday
Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
87498
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
ENTP | Enterovirus PCR, P | 97151-5 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
56068 | Enterovirus PCR, P | 97151-5 |
Reject Due To
Gross hemolysis | Reject |
Method Name
Real-Time Polymerase Chain Reaction (PCR)/RNA Probe Hybridization
Method Description
For this real-time reverse-transcription laboratory-developed polymerase chain reaction (PCR) assay, viral nucleic acid is extracted from specimens using the MagNA Pure automated instrument (Roche Applied Science), followed by amplification and detection on the Roche LightCycler 2.0 instrument. This PCR assay has been optimized to detect a target sequence in the polyprotein region. Primers amplify a 193 base-pair product.
Enterovirus genomic RNA is first transcribed to complementary DNA (cDNA) by reverse transcriptase, followed by amplification of the cDNA product. The LightCycler instrument can rapidly (30-40 minutes) detect amplicon development through stringent air-controlled temperature cycling in capillary cuvettes. The detection of amplified products is based on the fluorescence resonance energy transfer (FRET) principle. For FRET product detection, a hybridization probe with a donor fluorophore, fluorescein, on the 3'-end is excited by an external light source and emits light that is absorbed by a second hybridization probe with an acceptor fluorophore, LC-Red 640, at the 5'-end. The acceptor fluorophore then emits light of a different wavelength that can be measured with a signal proportional to the amount of specific PCR product. FRET (with subsequent production of a detectable fluorescent signal) only occurs when the probes have specifically annealed to the target sequence of the amplicon.
Melting curve analysis is performed following PCR amplification and is the detection phase of the assay, since it offers greater sensitivity than the amplification phase and maintains high specificity.
The melting phase of the assay occurs as follows:
Starting at 45° C, which allows the probes to bind to the amplified product, the temperature in the thermal chamber is then slowly raised to 80° C and the fluorescence measured at frequent intervals to determine the point where half of the fluorescence is lost as the probes are denatured (ie, "melt") off of the target. This is called the melting temperature (Tm) of that virus. Analysis of the PCR amplification and probe melting curves is accomplished through the use of LightCycler software.(Bernard PS, Reiser A, Pritham GH: Mutation detection by fluorescent hybridization probe melting curves. In: Meuer S, Wittwer C, Nakagawara K, eds. Rapid Cycle Real-Time PCR Methods and Applications. Springer; 2012:11-20)