Test Code Epic LAB181 Beta-2-Microglobulin (Beta-2-M), Spinal Fluid

Additional Codes

SQ code SFB2M

Mayo code B2MC

Reporting Name

Beta-2-Microglobulin, CSF

Specimen Type

CSF

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Useful For

Evaluation of central nervous system inflammation and B-cell proliferative diseases


Specimen Required


Container/Tube: Sterile vial

Specimen Volume: 1 mL

Collection Instructions: Send spinal fluid from collection vial 1.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
CSF Refrigerated (preferred) 28 days
  Frozen  28 days
  Ambient  14 days

Reference Values

0.70-1.80 mcg/mL

Day(s) Performed

Monday through Friday

Test Classification

This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

82232 

LOINC Code Information

Test ID Test Order Name Order LOINC Value
B2MC Beta-2-Microglobulin, CSF 1951-3

 

Result ID Test Result Name Result LOINC Value
B2MC Beta-2-Microglobulin, CSF 1951-3

Reject Due To

Gross hemolysis OK

Method Name

Nephelometry

Method Description

In this Siemens Nephelometer II method, the light scattered onto the antigen-antibody complexes is measured. The intensity of the measured scattered light is proportional to the amount of antigen-antibody complexes in the sample under certain conditions. If the antibody volume is kept constant, the signal behaves proportionally to the antigen volume. A reference curve is generated by a standard with a known antigen content on which the scattered light signals of the samples can be evaluated and calculated as an antigen concentration. Antigen-antibody complexes are formed when a sample containing antigen and the corresponding antiserum are put into a cuvette. A light beam is generated with a light emitting diode, which is transmitted through the cuvette. The light is scattered onto the immuno-complexes that are present. Antigen and antibody are mixed in the initial measurement, but no complex is formed yet. An antigen-antibody complex is formed in the final measurement. The result is calculated by subtracting the value of the final measurement from the initial measurement. The distribution of intensity of the scattered light depends on the ratio of the particle size of the antigen-antibody complexes to the radiated wavelength.(Instruction manual: Siemens Nephelometer II, Siemens, Inc.; Version 3, 2008; Addendum to the Instruction Manual 2.3, 08/2017)

Report Available

1 to 3 days