Test Code Epic LAB1050 Manganese, Blood
Additional Codes
SQ code MANGM
Mayo code MNB
Reporting Name
Manganese, BSpecimen Type
Whole bloodPerforming Laboratory
Mayo Clinic Laboratories in RochesterUseful For
Evaluation of central nervous system symptoms, similar to Parkinson disease, in manganese (Mn) miners and processors
Characterization of liver cirrhosis
Therapeutic monitoring in treatment of cirrhosis, parenteral nutrition-related Mn toxicity, and environmental exposure to Mn
Specimen Required
Patient Preparation: High concentrations of gadolinium and iodine are known to interfere with most metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Container/Tube: Royal blue top (EDTA) Vacutainer plastic trace element blood collection tube
Specimen Volume: 0.3 mL
Collection Instructions:
1. See Metals Analysis Specimen Collection and Transport for complete instructions.
2. Send whole blood specimen in original tube. Do not aliquot.
Specimen Minimum Volume
0.2 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Whole blood | Refrigerated (preferred) | 28 days | |
Ambient | 28 days | ||
Frozen | 28 days |
Special Instructions
Reference Values
4.7-18.3 ng/mL
Day(s) Performed
Tuesday
Test Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
83785
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
MNB | Manganese, B | 5681-2 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
89120 | Manganese, B | 5681-2 |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Triple-Quadrupole Inductively Coupled Plasma-Mass Spectrometry (ICP-MS/MS)
Method Description
The metal of interest is analyzed by triple-quadrupole inductively coupled plasma mass spectrometry.(Unpublished Mayo method)